Abstract

A new type of compact supersonic jet/resonance-enhanced multiphoton ionization/time-of-flight mass spectrometer is described. The analytical instrument, consisting of a single turbo molecular pump equipped with a rotary pump, was maintained at < 2 x 10(-3) Pa when a 0.3-atm sample was injected into a vacuum at 10-Hz using a 200-micros pulse valve. The diameters of the extraction and ground skimmer electrodes were expanded to 30 mm in order to avoid strong focusing and defocusing of the ion, and the optimum conditions for the system were investigated. The mass spectrometer functioned as expected: (1) no defocusing of the ion beam was observed even when the potential of the einzel lens was adjusted to zero; (2) the direction of the ion beam to an assembly of microchannel plates deviated in the expected manner when the potential of the defection electrode was changed from 0 to 30 V.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.