Abstract

A colorimetric method was developed for the determination of uranium in groundwater. The detection limit for the method is approximately 25 μg/L uranium, which is below the maximum contaminant level for uranium in drinking water, 30 μg/L. The method is rapid and requires little technical training to conduct, allowing it to be used by consumers, in the laboratory, or in the field. The two-step technique involves preconcentrating uranium using a U/TEVA-2 extraction chromatographic resin followed by complexation with a pyridylazo indicator dye, 2-(5-bromo-2-pyridylazo)-5-diethylaminophenol. If color change is visible to the eye, the concentration of uranium in groundwater is above the detection limit. Preconcentration using U/TEVA-2 also serves to eliminate metals that may interfere with the quantification of uranium.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.