Abstract
A rapid and sensitive UPLC–MS/MS method was developed and validated for determination of daclatasvir (DAC) in human plasma using sofosbuvir (SOF) as an internal standard (IS). The Xevo TQD LC–MS/MS was operated under the multiple-reaction monitoring mode using electrospray ionization. Precipitation with acetonitrile was used in sample preparation. The prepared samples were chromatographed on Acquity UPLC HSS C18 (50×2.1mm, 1.8μm) column by pumping 10mM ammonium formate (pH 3.5) and acetonitrile in an isocratic mode at a flow rate of 0.30ml/min. Method validation was performed as per the FDA guidelines and the standard curves were found to be linear in the range of 5–4000ng/ml for DAC. The intra-day and inter-day precision and accuracy results were within the acceptable limits. A very short run time of 1.2min made it possible to analyze more than 500 human plasma samples per day. The wider range of quantification of DAC allowed the applicability of the developed method for its determination in a bioequivalence study in human volunteers.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.