Abstract

The aim of the present work was to develop and validate a simple, economical, efficient and method for the analysis of Bosentan in pharmaceutical dosage forms by reverse phase high pressure liquid chromatography. A Welchrom C-18 reverse phase column (4.6 x 250 mm, 5 μm). With mobile phase containing 10mM Phosphate Buffer (pH 6.0): Acetonitrile (50: 50, v/v) is used and eluents were monitored at 226nm. The retention time of Bosentan was 3.687min. The method showed a good linearity in the concentration range of 15–27 μg/mL. The validation characteristics included specificity, linearity and limit of detection, limit of quantification, precision, robustness and stability. Validation acceptance criteria were met in all cases. The percent recoveries ranged between 90.7–100 and RSD 0.28%. The method could be successfully used for the analysis of Bosentan in pharmaceutical dosage forms

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.