Abstract

A sensitive, reproducible, robust, high-throughput ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) method was developed and validated for the simultaneous quantification of fexofenadine and olmesartan in human serum. Samples (50 µL) undergo protein precipitation prior to UPLC-MS/MS analysis. The analytes were separated using an Acquity BEH C18 column (2.1 mm × 50 mm, 1.7 µm) at a flow rate of 0.5 mL/min using a gradient elution with a total run time of 4 min. The analytes were detected in positive ion mode and selected reaction monitoring (SRM) was used for quantitation. The standard curve concentration range was 1.0–500.0 ng/mL for both analytes and each analyte showed excellent linearity with correlation coefficients (R2 > 0.99). The intra- and inter-day accuracy and precision were ±15% for each analyte, and excellent recovery was demonstrated (93–98%) for both analytes. The method is well suited for high-throughput quantitative determination of fexofenadine and olmesartan simultaneously and was successfully applied to an in vivo pharmacokinetic and transporter phenotyping study in humans.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.