Abstract
A highly sensitive, specific and high-throughput LC-ESI-MS/MS method in the positive mode has been developed and validated for the quantitation of Orteronel® (TAK-700) in rat plasma using YM-55208 as an internal standard. The assay procedure involves extraction of Orteronel and internal standard from rat plasma with liquid-liquid extraction method. Chromatographic separation was achieved using an isocratic mobile phase at a flow rate of 0.6 ml/min on an Atlantis dC18 column with a total run time of 2.5 min. The MS/MS ion transitions monitored were m/z 308.4→95.0 for Orteronel and m/z 262.3→194.2 for IS. Method validation was performed as per US FDA guidelines. The LLOQ achieved was 0.42 ng/ml and the linearity range extended from 0.42 to 814 ng/ml. The results met the acceptance criteria. The validated method was successfully applied to characterize the pharmacokinetic parameters of Orteronel in rat plasma.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.