A highly sensitive, specific and high-throughput LC-ESI-MS/MS method in the positive mode has been developed and validated for the quantitation of Orteronel® (TAK-700) in rat plasma using YM-55208 as an internal standard. The assay procedure involves extraction of Orteronel and internal standard from rat plasma with liquid-liquid extraction method. Chromatographic separation was achieved using an isocratic mobile phase at a flow rate of 0.6 ml/min on an Atlantis dC18 column with a total run time of 2.5 min. The MS/MS ion transitions monitored were m/z 308.4→95.0 for Orteronel and m/z 262.3→194.2 for IS. Method validation was performed as per US FDA guidelines. The LLOQ achieved was 0.42 ng/ml and the linearity range extended from 0.42 to 814 ng/ml. The results met the acceptance criteria. The validated method was successfully applied to characterize the pharmacokinetic parameters of Orteronel in rat plasma.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.