Abstract

Objectives The aim of this work was to implement a fast, accurate and simple method to quantify plasma ADMA and SDMA, in a run time suitable for routine analysis. Design and methods We developed and validated a hydrophilic interaction chromatographic method coupled to tandem mass spectrometry (HILIC–MS/MS) for separation and simultaneous quantification of Arginine (Arg) and its dimethylarginines, ADMA and SDMA, with a short run time (less than 5 min) using a small volume of human plasma (0.02 mL). Results Correlation coefficients ( r) of the calibration curves ranged from 0.9926 to 0.9984. Within-day and between-day imprecision (CV%) and inaccuracy (%), carry-over and recovery were also evaluated for validation. Preliminary data of Arg, ADMA and SDMA from 30 apparently healthy subjects and type 2 diabetic patients ( n = 33) with and without kidney dysfunction were calculated and some statistical differences occurred among them ( p < 0.05). Conclusions Data from calibration curves and quality controls reveal that the method is accurate and precise. Healthy subjects and diabetic patients' values are in agreement with those reported in other studies.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.