Abstract

A simple and highly sensitive CE-UV method was applied in the determination of l-ctrulline, which was developed from an oral formulation for pediatric use. The novel method was based on the analysis of l-citrulline for direct ultraviolet detection at 198nm. The BGE consisted of 10 mM sodium tetraborate and 50 mM SDS at pH 9, and the electrophoretic parameters were optimized. The method was validated in terms of specificity, linearity, LOD, LOQ, precision, accuracy, and robustness. The LOD and LOQ obtained were 1.36 and 4.54 μg/mL, respectively. In addition, the method offers higher sensitivity and specificity compared with the results obtained from HPLC method using UV-detectors, in which l-citrulline needs to be derivatizated. Furthermore, low cost and simplicity of the system allowed the rapid and simple quantitation of l-citrulline in the oral formulation for quality control and stability indicated method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.