Abstract

The calibration facility with the multi-nozzle calibrator was developed for the calibration of flow meters to be used with high-pressure, high-flow-rate hydrogen gas. The critical nozzles installed in the multi-nozzle calibrator were calibrated with traceability to the national standard. The relative standard uncertainty of the mass flow rates produced from the calibration facility is 0.09% when the flow rate is between 150g/min and 550g/min. In this study, the Coriolis flow meter was calibrated for a pressure range of 15–35MPa. The relative standard uncertainty of the flow rates obtained from the Coriolis flow meter was 0.44% for the case of the worst fluctuations in the output of the flow meter; based on the calibration curve, this is 0.91%. The present result shows that there is a maximum 3% difference between the output of the Coriolis flow meter and the mass flow rates of the multi-nozzle calibrator, even though the Coriolis flow meter was calibrated using water. Therefore, for the development of a calibration facility that can calibrate a flow meter under the same conditions as those encountered in actual use, it will be important to develop a new flow meter.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.