Abstract

A plasticised poly(vinyl chloride) optode membrane incorporated with a calix[6]arene hexaester, a H+-selective chromoionophore (3,3′,5,5′-tetramethyl-N,N-dibenzylbenzidine, a novel synthetic cationic dye) and a lipophilic potassium tetrakis(4-chlorophenyl)borate was used as a sensing device for the indirect optical determination of potassium ions. It exhibited a reversible response to the potassium ion in 0.05 mol/l HCl buffer media in the concentration range from 1 × 10− 6 to 1 × 10− 2 mol/l. The linear range was from 1.53 × 10− 5 to 3.20 × 10− 3 mol/l. The proposed optode sensor exhibited a fast response of less than 1 min, good repeatability (n = 7, RSD = 3.62%) at 5 × 10− 5 mol/l, and long-term stability with 92% of its initial sensitivity after 1 month of storage. The selectivity of the potassium-selective membrane allows its application for the detection of the potassium concentration in real sample analysis. The result was satisfactory compared with atomic absorption spectrometry.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.