Abstract

Summary In this paper we describe the construction of a laboratory-made resonant apparatus for the modulated force thermomechanical analysis (mf-TMA) and estimate the uncertainty of the measurement of Young’s modulus of a porcelain sample. The porcelain sample was measured while increasing the temperature up to 1000 C/5 C/min. Thermodilatometry was performed to determine the actual dimensions of the sample in the same temperature regime. We assessed the influence of the repeatability of the directly measured values, the accuracy of the used measurers and di erent exceptions from the ideal experimental arrangement on the Young’s modulus measurement uncertainties of types A and B. We found that the majority of the total uncertainty (80%) is determined at the room temperature. The remaining sources of uncertainty can be attributed to thermal expansion of the sample and inhomogeneity of the temperature field. The relative expanded uncertainty of the Young’s modulus at the elevated temperatures was 0.9%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.