Abstract
In the previously described flow-injection method for the spectrophotometric determination of water with the Karl Fischer reagent, problems can arise from the rather large variation between the calibration curves for different types of samples. Different combinations of flow-injection arrangements and detector types are described here, to establish a system which levels out this undesirable spreading between the calibration curves. One necessary condition for attaining a low spread is shown to be the use of a spectrophotometric detector especially constructed to minimize refractive index effects. The best results, including samples with large differences in refractive indices and viscosities, were obtained by a combination of peak area measurements and the use of this detector. For example, the relative mean deviation for a sample containing 0.1% water was about 2%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.