Abstract
We have developed an automated high-throughput assay for the determination of vitamin A (retinol), ergocalciferol (25-OH D2), cholecalciferol (25-OH D3) and vitamin E (α-tocopherol) in a small volume of human plasma. Sample preparation involved mixing 50 μL of plasma with 100 μL of ethanol containing isotope-labelled internal standards, followed by mixing with isooctane/chloroform (3:1, 300 μL). The organic phase was evaporated, and the sample reconstituted in 50 μL methanol. The analysis was performed using reversed-phase liquid chromatography with a gradient mobile phase containing water, methanol and ammonium formate. Chromatographic run-time was 5 min, and positive mode electrospray tandem mass spectrometry (MS/MS) was used for detection. The limits of detection were 0.10 μM for all-trans retinol and 3.3 nM for 25-OH D2 and 25-OH D3. Recoveries were 91.9-105.0%, and within- and between-day coefficients of variance (CVs) 2.4-5.3 and 3.1-8.2, respectively. The assay is presently being used in large-scale studies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.