Abstract

A liquid chromatography-electrospray ionization tandem mass spectrometry (HPLC-ESI-MS/MS) method for the quantitation of triamcinolone in human plasma after nasal spray application was developed and validated. Betamethasone was used as internal standard (IS). The analytes were extracted by a liquid-liquid procedure and separated on a Zorbax Eclipse XDB C(18) column with a mobile phase composed of 2 mM aqueous ammonium acetate pH 3.2 and acetonitrile (55:45). Selected reaction monitoring was performed using the transitions m/z 435 → 415 and m/z 393 → 373 to quantify triamcinolone acetonide and betamethasone, respectively. Calibration curve was constructed over the range of 20-2000 pg/ml for triamcinolone acetonide. The lower limit of quantitation was 20 pg/ml. The mean RSD values were 4.6% and 5.7% for the intra-run and inter-run precision, respectively. The mean accuracy value was 98.5% and a recovery rate corresponding to 97.5% was achieved. No matrix effect was detected in the samples. The validated method was successfully applied to determine the plasma concentrations of triamcinolone acetonide in healthy volunteers, in a pharmacokinetic study with nasal spray formulation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.