Abstract

Trace impurities in high-purity aluminium oxide were determined by high resolution inductively coupled plasma mass spectrometry (HR-ICP-MS). Samples were decomposed by sulfuric acid in PTFE pressure vessels. Most of the spectral interferences could be avoided by measuring in the high resolution mode (maximum mass resolution ( R), 5000). The matrix effects due to the presence of excess sulfuric acid and Al were evaluated. Since the effect of high Al concentrations on the peaks of several internal standard elements was not similar to the effect on the analyte elements, the standard addition method was employed for quantitative analysis. The detection limits in the solid samples were in the range of 0.01–4 μg g −1. Results for the determination of B, Na, Mg, Si, Ca, Ti, V, Cr, Mn, Fe, Co, Ni, Ga, Ge, Mo, Sn, Sb, Ba, La, Ce, W, Pt, Pb and U in several high-purity aluminium oxide powders were presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.