Abstract
An electrothermal atomic absorption spectrometric based method is described for the determination of total mercury in human whole blood. The blood is first treated with dilute hydrochloric acid to free the mercury species bound to protein thiol groups, then a pH buffer and a complexing agent (diethyldithiocarbamate) are added and the mercury species are extracted into toluene. In initial experiments, total mercury was determined in the toluene extract by electrothermal atomic absorption spectrometry. Recoveries of methyl-, ethyl- and phenylmercury chloride and mercury(II) chloride added to blood were better than 95% at concentrations up to about 30 µg dm–3. However, problems with excessive background absorbance were frequently observed, which could be overcome by back-extraction into dilute hydrochloric acid; this modification being adopted in the proposed method. The accuracy of the method was confirmed by the determination of total mercury in three lyophilized, human whole blood samples (Seronorm), the results demonstrating good agreement with the reference values.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.