Abstract

The aim of this work was to determine the concentrations and isotopic compositions of plutonium and uranium in environmental samples for safeguards purposes. An analytical method was developed with a plutonium and uranium separation procedure based on extraction chromatography (using 2 mL TEVA and UTEVA columns) and detection with a quadrupole ICP-MS applying an ultra-sonic nebulizer coupled with a membrane desolvation system. Starting from blank swipes, the background equivalent concentration (BEC) was 8 fg for 239Pu and 1 ng 238U. The method was successfully applied to certified reference materials as well as to round robin samples obtained in the framework of the inter-laboratory exercise program, promoted by the Brazilian–Argentine Agency for Accounting and Control of Nuclear Materials (ABACC), together with the US Department of Energy (USDOE). After the introduction of an additional ion-exchange separation step, the methodology was applied to the IAEA-384 sediment reference sample with precise and accurate total plutonium and uranium, 240Pu/ 239Pu, 241Pu/ 239Pu, 234U/ 238U and 235U/ 238U atomic ratio results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.