Abstract

The radial dependence of the diffuse reflectance from a turbid medium that is due to a point source is basically influenced by the absorption and reduced scattering coefficients. A system consisting of a HeNe laser source and a CCD camera is described for making remote measurements of the spatially resolved diffuse reflectance. Liquid tissue phantoms were made of Intralipid and trypan blue to validate the experimental setup. We show that for the correct determination of the optical properties of the tissue phantoms, the point-spread function (PSF) of the camera system has to be considered. Convolving the PSF with a solution of the diffusion equation, the absorption and reduced scattering coefficients of the tissue phantoms could be determined with an average error of 8% in the absorption coefficient and 4% in the reduced scattering coefficient, whereas without convolution, the errors were considerably larger, especially for large optical parameters.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.