Abstract
Sample heating due to high incident laser power in Raman experiments on (industrial) polymers and catalysts can easily lead to unwanted changes in morphology or desorption of surface species. Therefore temperature of the sample at the laser spot was studied by determining Stokes and anti-Stokes Raman intensities. Temperature can normally be determined within an accuracy of about 5 K, although for samples having nothing but Raman bands above 1000 cm−1, the anti-Stokes lines are very weak and accuracy of temperature determination decreased drastically. For most of the samples studied in the macro-Raman setup, powers up to 250 m\V could be used without measurable heating. In the micro-Raman setup heating started earlier due to higher focusing.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.