Abstract

Abstract A new amperometric titration method was developed for quantitative determination of ferrate(VI) (FeVIO42−) in the 7.06×10−5−5.73×10−3 M concentration range. Chromium(III) hydroxide solution was used as the titrant. The diffusion current (Id) had a linear relationship with the concentration of ferrate(VI) and slopes were dependent on the concentration of NaOH. The amperometric titration could detect a lower concentration of ferrate(VI) than could potentiometric and colorimetric titrations. The method was applied successfully to determine concentrations of ferrate(VI), generated electrochemically, in strong alkaline solutions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.