Abstract
A simple and sensitive method for the determination of steroids using micellar liquid chromatography is described. The steroids, including hydroxycorticosterone. corticosterone, northisterone, testosterone, mexdroprogesterone acetate and progesterone, were separated by reversed-phase using a micelles mobile phase following UV detection at 245 nm. The parameters affecting retention of the test solutes such as the concentration of sodium dodecyl sulfate (SDS) and n-butanol-1 in the mobile phase were investigated. It was found that the retention of the solutes was dependent on the composition of mobile phase. The linear calibration plots range from 0.1 to 10 μg ml−1 in mobile phase containing 5.0 × 10−2 mol l−1 SDS/9 % n-butanol-1 at pH 6.0, and the detection limit in order of 0.1 μg ml−1 was obtained. The proposed method was used for the determination of steroids in urine using direct injection of samples without previous treatment.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.