Abstract

The development of an amperometric sensor for the determination of reduced glutathione (GSH) is described. The sensor is based on tetrathiafulvalene–tetracyanoquinodimethane (TTF–TCNQ) incorporated into the graphite powder/Nujol oil matrix. The electrooxidation of GSH was monitored amperometrically at 200 mV versus SCE (saturated calomel electrode). The amperometric response of the sensor was linearly proportional to the GSH concentration between 20 and 300 μmol l −1, in 0.1 mol l −1 phosphate buffer (pH 8.0), containing 0.1 mol l −1 KCl and 0.5 mmol l −1 Na 2H 2EDTA, as supporting electrolyte. The detection limit, considering signal/noise ratio equal three, was 4.2 μmol l −1 for GSH and the repeatability obtained as relative standard deviation was of 5.1% for a series of 10 successive measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.