Abstract

A high-performance liquid chromatographic (HPLC) method is described for the determination of ractopamine (LY031537) in monkey plasma and swine serum. Plasma or serum (0.5 ml) was diluted with phosphate buffer pH 7.0. Ractopamine was isolated from the plasma matrix using ion exchange on a polymeric carboxylic acid solid-phase extraction cartridge followed by partitioning with ethyl acetate. An isocratic HPLC method using electrochemical detection at +700 mV was used to separate and measure ractopamine in the purified extract in 6.5 min of run time. Standard area response was linear with respect to concentration of ractopamine over the range of 0.5 to 40 ng/ml. Validation data were collected using rhesus monkey plasma and swine serum. The method precision and accuracy were evaluated in the range 1.0 to 20 ng/ml using fortified samples of monkey plasma. The method limit of quantitation was estimated at 2 ng/ml as determined in monkey plasma.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.