Abstract
Determination of two pyridinium ionic liquid cations (1-ethyl pyridinium cation and 1-butyl pyridinium cation) by ion chromatography using a carboxylic acid based cation-exchange column and direct conductivity detection was carried out. The optimized chromatographic conditions were 8.0 m mol L−1 methanesulfonic acid +5% acetonitrile (v/v) as eluent, flow rate of 1.0 mL min−1, and column temperature of 45°C. Detection limits (S/N = 3) for the two cations were 1.93 and 4.68 mg L−1, respectively. Relative standard deviations (n = 5) for peak areas were less than 0.8%. This simple method was successfully applied to determine pyridinium ionic liquids in spiked surface water.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Liquid Chromatography & Related Technologies
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.