Abstract
This radial compression liquid-chromatographic assay for propranolol in plasma is rapid, reproducible, and suitable for use in routine monitoring. A 10-micron particle, 8 mm X 10 cm CN cartridge is used in conjunction with a radial compression separation system. The mobile phase is monobasic sodium phosphate (pH 3) solution/methanol/acetonitrile (760/84/156 by vol), the flow rate 6 mL/min. Propranolol was detected by use of a spectrofluorometer equipped with a 20-microL flow-through cell, at excitation and emission wavelengths of 250 and 336 nm. The retention times for propranolol and metoprolol (the internal standard) are 3.13 and 1.42 min, respectively. A one-step extraction with chloroform yields "clean" chromatograms, with greater than 90% of the drug being analytically accounted for. Under these conditions, results are precise and accurate. Currently we are using this method to monitor propranolol in hypertensive neonates. Data on changes in the concentrations of propranolol in plasma with time are presented for one such patient.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.