Abstract
Thirteen laboratories participated in a collaborative study to validate a colorimetric assay for determining neutral lactase activity in industrial enzyme preparations. Each laboratory received 5 duplicate samples with activity levels of 2000 and 5000 neutral lactase units provided by 4 commercial suppliers. Two laboratories did not return results. Method performance was calculated according to AOAC guidelines. From the 11 remaining laboratories, 3 were excluded from statistical analysis because of invalid data determined during initial review by Youden pair, value versus laboratory. Repeatability relative standard deviation (RSDr) values ranged from 3.20 to 8.62%, and reproducibility relative standard deviation (RSDR) values ranged from 8.77 to 16.35%. With outliers excluded, RSDr values ranged from 2.94 to 5.01%, and RSDR values ranged from 7.50 to 13.84%. The colorimetric enzymatic method for determining neutral lactase activity in industrial enzyme preparations has been adopted first action by AOAC INTERNATIONAL.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.