Abstract
A procedure was developed for determination of Np using HR Sequential ICP-AES. Mutual interference and background free spectral lines of Np at 295.66, 297.43, 297.49, 302.64 nm were identified. The samples of higher concentration (10–20 µg/mL) of Np had been analyzed. Analytical values of 10–20 µg/mL samples were found to be in good agreement with the expected amounts. The concentration of Np between 2 and 4 μg/mL slightly affected as compared to higher concentration. In U/Pu, Np-295.66, 297.43 nm lines were monitored which showed noticeable enhancement of their intensities. The repetitive analyses showed that the precision was better than 2% RSD.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.