Abstract

Abstract An investigation was undertaken to determine whether commercially available mid infrared filter instrumentation (i.e., Multispec MK1) could be used to determine moisture in a manner similar to the AOAC near infrared method. Using a specially designed IR scanning spectrophotometer capable of working with aqueous systems, functional sample (6.02 jtm) and reference (5.56 iim) wavelengths were determined for quantitating water-methanol mixtures, and a standard curve was derived. Skim milk powder, bread crumbs, and flour were mixed with dry methanol to extract moisture into the solvent; moisture was quantitated relative to 3 AOAC methods. The results indicated that the wavelength pair selected worked well and had the potential to determine moisture content of low moisture food products, given that appropriate moisture extraction protocols were derived. This work indicates that the capabilities of commercial filter-based mid infrared instruments could be expanded to include moisture analyses if such instruments were equipped with the appropriate interference filters to carry out this determination.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.