Abstract
Electrothermal vaporization isotope dilution inductively coupled plasma mass spectrometry (ETV-ID-ICP-MS) was applied to the determination of mercury in urine samples. The isotope ratio of mercury in each analytical run was calculated from the peak area of each isotope. Various modifiers were tested in order to obtain the greatest sensitivity for mercury. After preliminary studies, a mixed modifier of palladium, Mg(NO 3 ) 2 and 2% HCl was used. The ETV-ID-ICP-MS method was applied to the determination of mercury in NIST SRM 2670 Toxic Metals in Freeze-Dried Urine and several fresh urine samples collected from graduate students. The results obtained for NIST SRM 2670 agreed satisfactorily with the certified values. The results for other samples analysed by isotope dilution and by the method of standard additions agreed satisfactorily. The detection limit of mercury, calculated from the standard additions calibration plot, was about 0.02 ng ml –1 with the ETV-ICP-MS method. The precision between sample replicates was better than 8% for all determinations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.