Abstract
The requirement for primary-level calibration of complex forms, e.g., gear profiles, has led to development of a method for quantifying the measurement uncertainty on the artifact being calibrated that does not rely on a transfer comparison. This method, developed jointly by National Institute of Standards and Technology (NIST) and Oak Ridge Metrology Center (ORMC) personnel, consists of breaking down the measurement method into simpler components and quantifying these components using generally accepted artifacts with low uncertainties (gage blocks, ring gages, spheres, etc.). Once quantified, these components are added vectorially according to NIST Technical Note 1297, 1994 Edition. Verification of this method can be accomplished by intercomparisons with other methods and laboratories.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.