Abstract

A combined method for macroconcentrations of platinum determination in chloroplatinic acid and industrial concentrate obtained during processing of some types of electronic engineering scrap that is based on a combination of gravimetric precipitation of platinum by ammonium chloride with subsequent determination of a residual concentration of platinum in a filtrate using the ISP-AES method is developed. The errors at each stage of analysis, as well as the total error, are estimated. It is shown that the accuracy of macro-concentrations of platinum determination using the combined method is not inferior to the accuracy of classical gravimetric analysis (the relative standard deviation does not exceed 0.15% for 15–40% wt concentrations of platinum). The combined method of analysis based on a combination of gravimetry and ISP-AES makes it possible to decrease labor expenditures and the time of analysis as compared to the classical one.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.