Abstract

To support pediatric study, a method to determine lumefantrine (LF) with small sample volume is needed. Matrix effect (ME) is a daunting issue in LF quantification in human plasma with LC-MS/MS. Here we report an LC-MS/MS method with a deuterated LF as the internal standard (IS). Plasma sample (25 µl) was acidified with 5% formic acid prior to extraction with ethyl acetate. The recovery was over 80%. The absolute ME was within the range of 100 ± 8% for both LF and the IS, but cumulative ME was observed via large variation of IS signal. The cumulative ME and ionization saturation were overcome with the co-eluting LF-D(9) as the IS. The linear range of calibration curve was 50-20,000 ng/ml. ME and ionization saturation was overcome with a deuterated IS. The method utilized a small sample volume, suitable for pediatric study with capillary tube blood collection method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.