Abstract
Implement a sensitive UHPLC method for the assay of ganciclovir in human plasma. We developed and validated a chromatographic method coupled to ultraviolet detection for quantification of ganciclovir, with a short run time using a small volume of human plasma. Comparison of system performance was made with respect to analysis time, efficiency and sensitivity. Correlation coefficients (r) of the calibration curves ranged from 0.999744 to 0.999784. Within-day and between-day imprecision and inaccuracy, specificity and recovery were also evaluated for validation. The method was precise and accurate and the retention time was 0.7 min. The calibration curves were linear between 0.5 and 30 μg/mL. There was a good correlation between HPLC and UHPLC techniques. We developed a method that is currently applied in a clinical study assessing GCV plasma concentration variability after ganciclovir and valganciclovir administration.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.