Abstract

Abstract The estrogen levels of amniotic fluid and maternal plasma may be altered in those cases with a fetus which presents Down syndrome. This paper describes the methodology for determination of free estriol in amniotic fluid. Fresh amniotic fluid was extracted using solid phase extraction techniques. Analtech C18 Spice cartridges were used to extract estriol (E3) prior to chromatography. After elution of the analyte E3 from the cartridge, the solvent was evaporated under nitrogen in a warm water bath at 60°C. The extract was reconstituted in methanol and E3 separated by high-performance liquid chromatography. A reversed phase C18 Partisil ODS-3 (4.6 mm × 25 cm) column was developed with acetonitrile—methanol (65:35 v/v) as mobile phase. Subambient temperature (−20°C) improved the resolution and aided separation from extraneous material. Quantitation followed intrapolation against a standard curve prepared under identical conditions. Measurement of amniotic fluid E3 may provide insight into the reporte...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.