Abstract

Two flow injection (Fl) methods, conventional and stopped-flow, are described for the determination of formaldehyde based on its reaction with pararosaniline in the presence of sulphite ion followed by spectrophotometric detection at 570 nm. The calibration graphs are linear over the ranges 2.0–50 and 0.2–10 µg ml–1 of formaldehyde for conventional and stopped-flow Fl, respectively. The coefficient of variation was 6.3% at the 0.6 µg ml–1 level in the stopped-flow mode, the detection limit being 0.1 µg ml–1 of formaldehyde. Sampling rates of 41 and 18 samples h–1 were achieved in the conventional and stopped-flow modes, respectively, this being a major advantage of these methods. The effect of several organic and inorganic species was studied and only low relative molecular mass aldehydes were found to interfere. The stopped-flow method was applied to the determination of formaldehyde in air in work environments and its accuracy was confirmed by comparing the results with those obtained using the standard chromotropic acid method.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.