Abstract

Several specific assays have been developed for the measurement of etoposide in biological fluids. As large samples are required for high sensitivity, these systems are not appropriate for a pediatric practice. In the present study, however, an improved method for the determination of serum levels of the anticancer drug etoposide was developed, using high-performance liquid chromatography with fixed-wavelength ultraviolet detection. Etoposide was extracted from serum using dichloromethane. The efficiency of extraction from serum was 85.7 +/- 7.7% for etoposide and 81.1 +/- 8.4% for diphenylhydantoin, the internal standard. The serum concentrations of etoposide were measured in 0.2-ml serum samples. The lower limit of detection was 50 ng/ml. Each measurement was completed within 5 min. The linear quantitation range for etoposide was 0.05-50 microg/ml. This assay presents an alternative method for routine measurement of serum levels of etoposide in the pediatric oncology setting.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.