Abstract
A method using gas chromatography/chemical ionization mass spectrometry (GC/CIMS) for the determination of daminozide residues in apples has been developed. Daminozide was separated from the sample matrix by water extraction and cation exchange, converted to the methyl ester by treatment with HCl-methanol, and determined by GC/CIMS using succinonitrile as an internal standard. The detection level was 0.05 ppm. Recoveries were 92-104% from apples spiked at the 0.05-0.5 ppm levels. Of the 25 apple samples analyzed, only 2 were positive for daminozide (1.04 and 0.32 ppm).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.