Abstract

Oxine/formaldehyde/resorcinol and oxine/formaldehyde/hydroquinone resins have been synthesized and their physicochemical properties studied. Conditions were optimized for the preconcentration of copper by batch extraction and column chromatography with the resins. A flow-injection analysis (FIA) manifold was constructed for the determination of copper at ng levels by preconcentration on microcolumns containing the resins, stripping, and atomic-absorption spectrometry. For batch preconcentration a pH of about 2.5–3 was optimal whereas in the FIA system a broader pH range (∼ 2–3.5) could be used. Separations of binary mixtures of Cu(II) with Ni(II) or PB(II) at U μg/ml level did not show any cross-contamination. In the FIA, a 2 cm long column and 2 ml/min flow-rate were adequate for quantitative uptake of copper; 50 μ1 of 0.1 M hydrochloric acid quantitatively eluted the copper.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.