Abstract

Two reversed-phase HPLC methods with UV detection to quantify celiprolol and oxprenolol in human plasma are described. The analytical methods for the determination of both drugs used the same reversed-phase HPLC column, mobile phase and extraction procedure. Linearity was obtained in the ranges 15.63–1000 and 25–800 ng/ml for celiprolol and oxprenolol, respectively. Intra-day and inter-day variation was lower than 14%. After validation of the methods, analytical error functions were established as S.D. (ng/ml)=3.096+0.041 C for celiprolol and S.D. (ng/ml)=8.906+8.075·10 −8 C 3 for oxprenolol.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.