Abstract
A method to determine catechin in aqueous solution by measuring chemiluminescence intensities using a stopped flow system has been studied. The lucigenin-hydrogen peroxide chemiluminescence reaction was chosen for the determination of catechin. Fe(II) ion was added to the chemiluminescence system to increase the sensitivity. The chemiluminescence intensity from the lucigenin system was increased by the addition of catechin. Effects of flow rates of reagent and sample and concentrations of lucigenin, hydrogen peroxide, Fe(II) ion and KOH were investigated. The calibration curve for catechin was linear over the range from 1.0x10(-6) to 1.0x10(-3) M and the detection limit was 3.0x10(-7) M under the optimal experimental conditions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.