Abstract

The direct analysis of solid samples using (GFAAS) graphite furnace atomic absorption spectrometry (GFAAS) has considerable potential as a method for the determination of trace metals when only small amounts of sample are available, when the highest accurracy is not necessarily required, or when sample dissolution would be difficult and/or time consuming. Whereas the complete dissolution of many plastics is very difficult and often time consuming, the analysis by solid sampling offers the potential for avoiding sample dissolution. Using the cup-in-tube technique, Cd, Cu, Mn and Rb have been determined in solid plastic samples using Zeeman-effect GFAAS. The analytical conditions used permitted the determinations to be performed using aqueous reference solutions. Acceptable agreement of the results with reference values was achieved.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.