Abstract

An inductively coupled plasma mass spectrometry (ICP-MS)-based immunoassay has been developed to quantify aflatoxin M1 (AFM1) at ultra-trace levels in milk samples. AFM1 detection is carried out by means of a competitive immunoassay using secondary biotinylated antibodies and streptavidin-conjugated Au nanoparticles. After acid addition, nanoparticles are decomposed and Au signal is registered by means of ICP-MS. Results demonstrate that, under optimum conditions, the limit of detection of the immunoassay (0.005μgkg-1) is low enough to quantify AFM1 according to current international policies (including the more restrictive European one). Method accuracy and precision was checked by analyzing an AFM1 certified reference material and different milk samples spiked with known amounts of AFM1. AFM1 recovery values range from 80% to 102% whereas inter-assay and intra-assay precision are lower than 15%. Finally, this immunoassay methodology affords a higher dynamic working range (0.012-2.5μgkg-1) than other immunoassay methodologies described in the literature.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.