Abstract
Activity-based probes (ABPs) are specific covalent inhibitors developed for different classes of enzymes. We have titrated a serine protease and a lipase with their specific ABPs and measured the extent of inhibition using nanoelectrospray mass spectrometry (nanoESI–MS). Because ABPs only interact with the active enzyme form, the approach allows to accurately measure the active enzyme concentration in solution. This is even possible in the presence of contaminants. The concentrations of the two enzymes were also investigated by UV spectroscopy, which appears to give higher concentrations than those measured with the active site titration method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.