Abstract
A novel analysis program to unfold alpha-particle energy spectra was introduced, demonstrated and validated using radiochemically processed test sources, which contained different amounts of 239Pu and 240Pu. A high-resolution alpha spectrometer was used for data acquisition. The software known as ADAM unfolds the spectra using nuclide-specific decay data as a constraint. The peaks can have different shapes and the software can also cope with the coincidences between alpha particles and electrons/photons. In the present paper, the 239Pu/240Pu activity ratios from alpha spectrometry agreed, within the stated uncertainties, with the reference values. Number of counts in the 239,240Pu peak group must be larger than 100 to obtain reliable values when using semiconductor detector of energy resolution FWHM=10.6keV.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.