Abstract

A fast analytical method for determination of hydroquinone in pharmaceutical formulations employing batch injection analysis (BIA) with amperometric detection using a boron-doped diamond electrode is described. The supporting electrolyte was a 0.1 mol L-1 H2SO4 solution (the single reagent used for analysis). The method showed good repeatability (RSD of 0.45%, n=20), wide linear range (from 10 to 2000 µmol L-1, R=0.9999), low detection limit (0.016 µmol L-1) and satisfactory recovery values (91-96%). Accuracy of the method was evaluated by comparative analyses using high-performance liquid-chromatography. The ability to replace the electronic pipette by disposable syringes (injection procedure) in BIA systems was also shown.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.