Abstract
Factors(integral potential,concentration of NaOH and flow rate) affecting detection of Glucosamine hydrochloride by ion chromatography with integrated pulsed amperometric detector were studied.As the integral potential increased,the detection signal first increased and then decreased,while the detection background and noise increased directly.The suitable integral potential should be 0.1 V.With the concentration of NaOH increased,the retention time descended sharply,while the integral area increased quickly before they are coming back into balance.The number of theoretical plates increased firstly and then decreased.The optimum concentration of NaOH was 50 mmol·L-1.As the flow rate increased,the retention time and peak area reduced.Thus,the flow rate of 0.6 mL·min-1was the suitable one.The transformation-quantity of Glucosamine hydrochloride using ED3000 detector was 1.27% by connecting the two integrated pulsed amperometric detectors,while the total amount was 2.79% by NJPAD detector.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.