Abstract

Modern implantable cardioverter/defibrillator (ICD) systems offer amultitude of algorithms to optimize performance in sensing and tachycardia detection even in difficult circumstances (e. g., ventricular tachycardia during supraventricular tachycardia, fine ventricular fibrillation with intermittent undersensing), to reliably discriminate sustained ventricular tachyarrhythmia from noise, nonsustained and supraventricular tachyarrhythmia, and to limit shock therapy only to those arrhythmias that definitely need to be treated by ashock. Adisadvantage of these multiple algorithms is the complexity of annotated tracings that makes it sometimes difficult to understand why the ICD did what it did. If atachycardia classification was wrong, it may be thus difficult to find the best way to reprogram the device to avoid another misclassification. This review explains in detail the algorithms used for tachycardia detection, discrimination, and prevention of inappropriate therapy in single- and dual-chamber ICDs manufactured by St. Jude Medical. Knowledge of these features may help to optimize ICD treatment in patients fitted with these devices.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.