Abstract
Cortical lesions have recently been a focus of multiple sclerosis (MS) MR research. In this study, we present a white matter signal attenuating sequence optimized for cortical lesion detection at 7 T. The feasibility of white matter attenuation (WHAT) for cortical lesion detection was determined by scanning eight patients (four relapsing/remitting MS, four secondary progressive MS) at 7 T. WHAT showed excellent gray matter–white matter contrast, and cortical lesions were hyperintense to the surrounding cortical gray matter, The sequence was then optimized for cortical lesion detection by determining the set of sequence parameters that produced the best gray matter–cortical lesion contrast in a 10-min scan. Despite the B1 inhomogeneities common at ultra-high field strengths, WHAT with an adiabatic inversion pulse showed good cortical lesion detection and would be a valuable component of clinical MS imaging protocols.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.