Electro capacitive cancer therapy (ECCT) system are constructed by six sub-systems, in this paper presented design and simulation for two sub-system in ECCT namely frequency generator and amplifier level converter. Design of two sub-system become first study for ECCT system can be applying to the IC technology. Design process begins with understand ECCT standard operation mode signal specification like input signal voltage, output signal voltage, form and frequency. Then using ring oscillator, negative clamper biased, two stage operational amplifier (op amp) circuits that estimated can produce those signal specification and layout them based on BSIM 3.1 MOSFET model using 300nm CMOS technology. Simulation result show that the system layout can produce 19Vpp, 100 kHz and asymmetric output signal from 5 volt DC input signal.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.