Abstract

This paper describes the salient points in establishing robustness of the cell-based potency assay, with reference to phase-specific qualification. The methodology was qualified in a thaw for use format utilizing standard parameters. At a later stage, the assay was further developed, and concomitant with requalification the robustness was assessed by fractional factorial design of experiment, the center point data being used for qualification. The methodology was qualified with a precision of 8.5% and mean bias of 3.6%. The robustness data generated a similar precision and accuracy/bias to the qualification data. Assay factors were optimized to be within the design space model of the design of experiment. The methodology was qualified and was robust with regard to assay parameters tested.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.